Veris Technologies
Pub. #OM MSP03-1
1
Mobile Sensor Platform
Table of Contents
General Information Section
P. 2 Warranty and Safety
P. 4 Instrumentinstallation and set-up
P. 5 File Managementdata downloading and transfer
P. 7 GPS trouble-shooting
P. 8 Instrument firmware update procedure
Soil EC Surveyor
P. 10 Installation and set-up
P. 12 Field Operations
P. 15 Maintenance and Lubrication
P. 21 Troubleshooting
pH Manager
P. 22 Installation and set-up
P. 25 Field Operations
p. 27 Adjusting sampling depth
p. 28 Other adjustments
p. 30 pH electrode calibration
p. 33 Collecting pH data
p. 39 Clean-up
P. 40 Maintenance and Lubrication
P. 45 Troubleshooting
Operating Instructions
Vue de la page 0
1 2 3 4 5 6 ... 60 61

Résumé du contenu

Page 1 - Mobile Sensor Platform

Veris Technologies Pub. #OM MSP03-1 1 Mobile Sensor Platform Table of Contents General Information Section P. 2 Warranty and Safety P. 4

Page 2

Veris Technologies Pub. #OM MSP03-1 10 Soil EC Surveyor Installation and Set-up If the unit has been crated and delivered via closed-van commercial

Page 3

Veris Technologies Pub. #OM MSP03-1 11 Fill gasoline engine and check engine and hydraulic oil levels. Before filling tank with gasoline, let en

Page 4

Veris Technologies Pub. #OM MSP03-1 12 • Escaping fluid under pressure can penetrate the skin causing serious injury. Avoid the hazard by relieving

Page 5 - File Management

Veris Technologies Pub. #OM MSP03-1 13 Signal Cable Attach the signal cable to the quick connect coupler at front of frame, and to Signal Port on b

Page 6

Veris Technologies Pub. #OM MSP03-1 14 Turn on Veris instrument. The first screen to appear contains the software version number and shows how much

Page 7 - GPS Troubleshooting

Veris Technologies Pub. #OM MSP03-1 15 SEE GENERAL INFORMATION SECTION FILE MANAGEMENT FOR INSTRUCTIONS ON DOWNLOADING DATA Maintenance and Lubrica

Page 8

Veris Technologies Pub. #OM MSP03-1 16 c) Continue to check each coulter electrode in succession, left to right. d) If any coulter electrode exhi

Page 9

Veris Technologies Pub. #OM MSP03-1 17 the commutators should be checked. This may be performed in the following manner: 1) Remove the

Page 10 - Soil EC Surveyor

Veris Technologies Pub. #OM MSP03-1 18 an upper and lower grease zerk. Due to the limited motion of the rockshaft, these should be lubricated on 40-

Page 11

Veris Technologies Pub. #OM MSP03-1 19 Wheel hubs -- On an annual basis, disassemble, clean, and properly repack the wheel hubs with su

Page 12

Veris Technologies Pub. #OM MSP03-1 2 VERIS Mobilized Sensor Platform (MSP) (Software Version MSP 1.02a) Warranty Veris Technologies warrants this

Page 13

Veris Technologies Pub. #OM MSP03-1 20 Seal Hub Nut Thrust Washer Bearing Swing arm

Page 14

Veris Technologies Pub. #OM MSP03-1 21 EC MODULE TROUBLESHOOTING Map doesn’t match known or expected soil types -- 1. Check electrical continuity us

Page 15 - Maintenance and Lubrication

Veris Technologies Pub. #OM MSP03-1 22 pH Manager Installation and Set-up If the unit has been crated and delivered via closed-van commercial freig

Page 16

Veris Technologies Pub. #OM MSP03-1 23 Before filling tank with gasoline, let engine cool. Gas vapors can ignite and an explosion can occur.

Page 17

Veris Technologies Pub. #OM MSP03-1 24 Connecting cables to External controller as shown below: Remove

Page 18

Veris Technologies Pub. #OM MSP03-1 25 Field Operations Tools required for Field Operation adjustments -3/16” allen wrench -adjustable wrench: min

Page 19

Veris Technologies Pub. #OM MSP03-1 26 2. Put external controller in Manual mode and raise and lower sampler assembly. Keep appendages clear of pin

Page 20

Veris Technologies Pub. #OM MSP03-1 27 2. Once unit is level, lower sampling mechanism completely, drive forward 10-20’ (3-6 m) to create soil cor

Page 21 - EC MODULE TROUBLESHOOTING

Veris Technologies Pub. #OM MSP03-1 28 Other Field Adjustments Once EC coulter depth and sampling depth are satisfactory, adjust other components in

Page 22

Veris Technologies Pub. #OM MSP03-1 29 4. Insert pH electrodes into electrode holder. Finger-tighten plastic screws. Install BNC cover on extern

Page 23

Veris Technologies Pub. #OM MSP03-1 3 Important! Read the following SAFETY PROCEDURES before operating the Veris MSP: • Read and understand all in

Page 24

Veris Technologies Pub. #OM MSP03-1 30 7. Prox sensor: The prox sensor communicates the position of the sampler assembly to the external controller

Page 25 - Field Operations

Veris Technologies Pub. #OM MSP03-1 31 7. You will be asked for the ID of the electrode connected to channel 1. Use the 1 and 2 keys to change the

Page 26

Veris Technologies Pub. #OM MSP03-1 32 If the readings are satisfactory, log pH 7 reading and press 1; if the readings are suspect, press 2 to return

Page 27

Veris Technologies Pub. #OM MSP03-1 33 Collecting pH data Set external controller to Automatic mode. Turn on Veris instrument. The first screen t

Page 28

Veris Technologies Pub. #OM MSP03-1 34 Maximum log time is the longest time in seconds the pH controller will wait for the pH readings to settle.

Page 29

Veris Technologies Pub. #OM MSP03-1 35 pH and EC (option 3): The display is showing the pH values from the pH electrodes, conductivity of the t

Page 30

Veris Technologies Pub. #OM MSP03-1 36 See the troubleshooting section if this occurs frequently. The Veris MSP pH Manager uses two electrodes for

Page 31

Veris Technologies Pub. #OM MSP03-1 37 Maximum log time is the longest time in seconds the pH controller will wait for the pH readings to settle. T

Page 32

Veris Technologies Pub. #OM MSP03-1 38 Exiting Data Acquisition and Extracting pH Data When you are ready to stop collecting pH data to the file, pr

Page 33

Veris Technologies Pub. #OM MSP03-1 39 the final settled smoothed soil pH readings (VPHE000.DAT). A formatting example of each file type follows: Ra

Page 34

Veris Technologies Pub. #OM MSP03-1 4 Instrument Installation and Set-up The Veris Instrument Kit includes the following items: Mount instrument

Page 35

Veris Technologies Pub. #OM MSP03-1 40 Maintenance and Lubrication • Pinch point hazard: to prevent injury, stand clear when raising or loweri

Page 36

Veris Technologies Pub. #OM MSP03-1 41 Wash System If wash water develops algae, flush and fill tanks with tap water; clean any algae or othe

Page 37

Veris Technologies Pub. #OM MSP03-1 42 Replace sampling trough liner, scraper cutting edge when wear is appa

Page 38

Veris Technologies Pub. #OM MSP03-1 43 Engine Sediment Cup Clean as Instructed in Honda Engines Manual. Engine Spark Plug Clean-Readjust as Instr

Page 39

Veris Technologies Pub. #OM MSP03-1 44 Change the filter cartridge whenever the hydraulic fluid is drained. Parker replacement cartridges (Veris P/N

Page 40

Veris Technologies Pub. #OM MSP03-1 45 pH Module Troubleshooting 1. Unit doesn’t cycle when “1” key is pressed (in Automatic Mode) -check to be sure

Page 41

Veris Technologies Pub. #OM MSP03-1 46 2. Functions Aren’t Working in Manual Mode -follow troubleshooting flow chart below and see related procedure

Page 42

Veris Technologies Pub. #OM MSP03-1 47 -check for moisture around BNC connectors and for proper installation and fit of the BNC cover over BNC conn

Page 43

Veris Technologies Pub. #OM MSP03-1 48 11. Electronic controller not functioning: remove metal plate covering external controller and inspect the LE

Page 44

Veris Technologies Pub. #OM MSP03-1 49 Procedure #1: Checking power inside External Controller The Veris pH module requires a minimum of 12 volts at

Page 45

Veris Technologies Pub. #OM MSP03-1 5 magneto, consider powering the Veris instrument with a 12-volt battery or converting to an alternator system.

Page 46

Veris Technologies Pub. #OM MSP03-1 50 Figure 2. External Controller circuit board (metal cover removed) 2. Turn off power to External Controller,

Page 47

Veris Technologies Pub. #OM MSP03-1 51 If voltage at power connector drops below 12 volts during this test: a) clean battery terminals b) check di

Page 48

Veris Technologies Pub. #OM MSP03-1 52 Figure 1. Flowchart for trouble-shooting sampler hydraulics. 1. If cylinder doesn’t move at all in manual

Page 49

Veris Technologies Pub. #OM MSP03-1 53 Figure 3. External Controller circuit board (metal cover removed) 3. Test voltage coming out of External

Page 50 - FUSE 2 (250

Veris Technologies Pub. #OM MSP03-1 54 Figure 5. Testing continuity of solenoid coil (at weather-pack connector on jumper-wire) 5

Page 51

Veris Technologies Pub. #OM MSP03-1 55 Procedure #3: Trouble-shooting Communication Problem between Instrument and External Controller Communicatio

Page 52

Veris Technologies Pub. #OM MSP03-1 56 Figure 2. Inspecting pins on communication cable 2. Testing continuity on communication cable. a

Page 53 - SOLENOID LED’s

Veris Technologies Pub. #OM MSP03-1 57 d) Check continuity from brown Molex connector to 9 pin serial connector; this tests continuity from the circ

Page 54

Veris Technologies Pub. #OM MSP03-1 58 5. Testing that the instrument is sending the proper command to the external controller. a) Connect the In

Page 55

Veris Technologies Pub. #OM MSP03-1 59 Procedure #4: Trouble-shooting Prox Sensor The prox sensor, located on the left of the sampler shoe parallel

Page 56 - 5 3 2

Veris Technologies Pub. #OM MSP03-1 6 About the Veris EC output file format Veris EC data is output as a 5 column ASCII text file. Column A: longitu

Page 57

Veris Technologies Pub. #OM MSP03-1 60 Figure 2. Prox sensor gap 3. Adjust prox sensor height i

Page 58

Veris Technologies Pub. #OM MSP03-1 61 Procedure # 5 Compact Flash card formatting Insert a Compact Flash card into a CF card reader which connected

Page 59

Veris Technologies Pub. #OM MSP03-1 7 This is the second window you will see after powering up the unit. Select 3 to download or delete files that

Page 60

Veris Technologies Pub. #OM MSP03-1 8 d) The program will then ask you for a phone number. Instead of entering a phone number, specify the proper s

Page 61 - Pub. #OM MSP03-1

Veris Technologies Pub. #OM MSP03-1 9 Running Utility Files Utility files can be run from the CompactFlash card to perform various tasks. Here are t