Veris-technologies pH Detector - Operating Instructions Manuel d'utilisateur Page 12

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 29
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 11
OM17-pH Detector
3-2
-Flush and fill tanks with tap water; clean any foreign matter out of tank using clean-out cap. Make
sure inlet is in sump (Figure 3), and strainer is attached to end of line (Figure 4).
Figure 3 Figure 4
Pump priming: In order to prime the pumps, loosen quick coupler and operate pump until water
flows (Figure 6).
Wash adjustment: This was set at the factory but may have slipped during shipping. Wash
brackets should be opposite each other, with jets as close as possible to electrode holder, and still
allow electrode holder movement. Jets should be 1/4” (6 mm) below electrodes. When wash jets
are properly aligned, overspray is minimized and water bubbles out top of empty electrode holder
as shown in Figure 5. To adjust wash jets, use 3/16” allen wrench to loosen set screws holding
wash jet blocks and rods (Figure 6).
Figure 5 Figure 6
set screws
Quick
coupler
Vue de la page 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 28 29

Commentaires sur ces manuels

Pas de commentaire