Veris-technologies MSP3 - Operating Instructions Manuel d'utilisateur Page 34

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 116
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 33
Pub. #OM17-MSP3
4-14
The software will prompt for the electrodes to be inserted into pH buffer 4 solution; Slide cup with
pH 4 buffer solution onto electrode holder. Press 1 to continue with calibration or 2 to exit. Tips:
Don’t overfill solution. Cup only needs enough solution to immerse electrode tip and face. Don’t
reuse solutions.
Figure 43
The software will read the electrodes for 10 seconds, displaying the output (as it counts seconds):
Figure 44
After 10 seconds, the software will display the final pH reading and offer the options to Accept pH 4
buffer readings; Redo pH 4 buffer readings; or Cancel pH electrode calibration. If the readings are
satisfactory, press Accept; if the readings are suspect, press Redo to return to re-do pH 4.
Figure 45
Vue de la page 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 115 116

Commentaires sur ces manuels

Pas de commentaire